Anatomic shoulder implant wear test PI-62

Normative References  



PI-62: Anatomic Shoulder System Wear Test

The test parameters for anatomic shoulder replacement systems are based on a wear study conducted by Wirth et al [1].

A modified EndoLab® hip joint simulator is used for the testing.

The following motions are applied simultaneously to the implant system: Abduction/adduction, flexion/extension, translation within the plane of abduction/adduction, translation within the plane of flexion/extension as well as forward/backward elevation. A constant load of 756 N (± 50 N) is applied. According to Wirth et al. the translation, resulting in rim loading during abduction/adduction is oriented at a 45° angle from the long axis of the glenoid component. The amount of translation required to generate rim loading during abduction/adduction depends on the radial mismatch between the glenoid component and the humeral component and will be adjusted individually. The wear test is conducted for 5 million cycles. The wear is determined gravimetically.  


[1] Michael A. Wirth, MD, Conrad Klotz, BS, MBa, Daren L. Deffenbaugh, MS, Don McNulty, MS, PE, Laura Richards, Beng, Joanne L. Tipper, PHD: Cross-linked glenoid prosthesis: A wear comparison to conventional glenoid prosthesis with wear particulate analysis. Journal of Shoulder and Elbow Surgery (2009) 18, 130-137  

Implant Testing - Anatomic shoulder implant wear test PI-62

CLOSE [X]
email problems with office 365
Dear clients & partners,

Since the end of January, we have heard from some clients who use Office 365 that they are experiencing trouble receiving our emails. Our emails seem to be classified as phishing attempts and, therefore, never reach their inboxes.

If you are also waiting for a reply, please call us at +49 8031 2313230!

We look forward to hearing from you again and hopefully re-establish full communication.

The EndoLab Team