Fatigue test baseplate/bearing fixation ASTM F2723

Normative References  



ASTM F2723: Standard Test Method for Evaluating Mobile Bearing Knee Tibial Baseplate/Bearing Resistance to Dynamic Disassociation

This test method describes a laboratory method for evaluating the potential for mobile bearing knee tibial baseplate/bearing disassociation under repeated forces.

The test described is applicable to any bicompartmental mobile bearing knee with a bearing retention mechanism. With modification, the test can be applied to a unicompartmental mobile bearing knee with a bearing retention mechanism.

A total of n=5 implants is tested at the worst case flexion angle regarding the retention mechanism.  


Implant Testing - Fatigue test baseplate/bearing fixation ASTM F2723