EndoLab - Implant Testing Services Implant Testing

Fatigue test baseplate/bearing fixation ASTM F2722

Normative References  



ASTM F2722: Standard Practice for Evaluating Mobile Bearing Knee Tibial Baseplate Rotational Stops

This practice covers a laboratory-based in vitro method for evaluating the mechanical performance of mobile bearing knee designs and their ability to maintain the integrity of the rotational stop feature.
The test is restricted to implants providing a rotational stop within 20° of rotation in internal or external direction. A maximum torque of 14 Nm is applied up to 220 000 cycles or implant failure.  


Implant Testing - Fatigue test baseplate/bearing fixation ASTM F2722

CLOSE [X]