EndoLab - Implant Testing Services Implant Testing

Knee implants
all procedures accredited

Knee Constraint Test - ASTM F1223Knee Constraint Test - ASTM F1223 Dislocation Mobile Bearing - ASTM F2724Dislocation Mobile Bearing - ASTM F2724 Knee Contact Pressure Test - PI-17Knee Contact Pressure Test - PI-17 Tibial Component Finite Element Analysis - ASTM F3334Tibial Component Finite Element Analysis - ASTM F3334 Tibial Tray Fatigue Test - ISO 14879-1/ ASTM F1800Tibial Tray Fatigue Test - ISO 14879-1/ ASTM F1800 Patella specification - ASTM F1672Patella specification - ASTM F1672 Knee surface analysis - ISO 7207-2Knee surface analysis - ISO 7207-2 Knee wear test - ISO 14243-1/3Knee wear test - ISO 14243-1/3 Fatigue Test under High Flexion - ASTM F2777Fatigue Test under High Flexion - ASTM F2777 Fatigue Test Baseplate/Bearing Fixation - ASTM F2723Fatigue Test Baseplate/Bearing Fixation - ASTM F2723 Fatigue Test Baseplate/Bearing Fixation under Torsional Loading - ASTM F2722Fatigue Test Baseplate/Bearing Fixation under Torsional Loading - ASTM F2722 Knee loading profiles - ASTM F3141Knee loading profiles - ASTM F3141 FEA knee femoral component - ASTM F3161FEA knee femoral component - ASTM F3161 Unicondylar tibial tray fatigue test - ASTM F3140Unicondylar tibial tray fatigue test - ASTM F3140 Knee Condyle Fatigue Test - ASTM F3210/ PI-73Knee Condyle Fatigue Test - ASTM F3210/ PI-73 Durability performance patellofemoral joint - ISO 14243-5Durability performance patellofemoral joint - ISO 14243-5 Stemmed femoral knee component fatigue test - PI-53Stemmed femoral knee component fatigue test - PI-53 Specific Requirements for Knee-joint Replacement Implants - ISO 21536Specific Requirements for Knee-joint Replacement Implants - ISO 21536 Test of Modular Knee Implants - ASTM F1814Test of Modular Knee Implants - ASTM F1814 Particular requirements for joint replacement implants - ISO 21534Particular requirements for joint replacement implants - ISO 21534 Knee Specification - ASTM F2083Knee Specification - ASTM F2083

Unicondylar tibial tray fatigue test - ASTM F3140

Normative References  



ASTM F3140: Standard Test Method for Cyclic Fatigue Testing of Metal Tibial Tray Components of Unicondylar Knee Joint Replacements

This test method covers a procedure for the fatigue testing of metallic tibial trays used in partial knee joint replacements. Testing is applicable for both the medial and lateral plateau of the tibia. For testing, the tibial tray is supported locally at the anterior and posterior side and loaded centrally by a spherical indenter. The suggested number of cycles is 10 million at a test frequency of 20 Hz or less.

 


Implant Testing - Unicondylar tibial tray fatigue test - ASTM F3140