EndoLab - Implant Testing Services Implant Testing

Vascular implants
 


Resistance to Perpendicularly-applied Load - ISO 25539

Normative References  



ISO 25539‐1: Cardiovascular implants - Endovascular devices - Part 1: Endovascular prostheses

ISO 25539‐2: Cardiovascular implants - Endovascular devices - Part 2: Vascular stents

The purpose of this test is to determine the force at which a pre-specified amount of permanent deformation (for balloon-expandable stents) or a pre-specified displacement (for self-expanding stents) occurs under a load applied perpendicular to the longitudinal axis of the stent.  


Implant Testing - Resistance to Perpendicularly-applied Load - ISO 25539