EndoLab - Implant Testing Services Implant Testing

Balloon inflation to deflation time ISO 25539

Normative References  



ISO 25539‐1: Cardiovascular implants - Endovascular devices - Part 1: Endovascular prostheses.

ISO 25539‐2: Cardiovascular implants - Endovascular devices - Part 2: Vascular stents.

ISO 10555-4: Intravascular catheters – Sterile and single-use catheters – Part 4: Balloon dilatation catheters.

The purpose of this test is to determine the time required to inflate the balloon to the maximum recommended inflation pressure, volume or diameter and to measure the time required to deflate the balloon. This test provides information that might be clinically useful for treatment planning (e.g. potential occlusion time).

EndoLab® uses a computerized pressure device which allows complex pressure processes and rates as well.  


Implant Testing - Balloon inflation to deflation time ISO 25539