EndoLab - Implant Testing Services Implant Testing

Durability performance patellofemoral joint PI-79 (ISO 14243-5)

Normative References  



PI-79: (ISO 14243-5) Implants for surgery – Wear of total knee prostheses – Part5: Durability performance of the patellofemoral joint – Loading and displacement parameters for testing machines and corresponding environmental conditions for test

Loading and displacement waveforms, which simulate the patellofemoral kinetics and kinematics by combining a squatting sub-cycle with two walking gait sub-cycles, are performed until delamination, cracking, or another identified failure mode occurs. If the implants do not demonstrate any failure modes by 220 thousand cycles, the test is stopped. The EndoLab® knee joint simulator is used for testing. Please download our video to see the simulator moving.  


Implant Testing - Durability performance patellofemoral joint PI-79 (ISO 14243-5)